Handicare USA, Inc. recalls ReTurn belt, wipeable, sizes Small through Extra Large, Plastic pouch, 1 item per pouch. Patient transfer aid.
Recalled May 20, 2015 · FDA recall Z-1613-2015 · Handicare USA
Hazard
Complaints were received that the heat-welded seam on the wipeable belt's strap tore when force was applied. · FDA Class II
Units
64
Sold at
US Distribution to the states of : FL, CA, NJ, NC and MD.
Description
Complaints were received that the heat-welded seam on the wipeable belt's strap tore when force was applied.
Product
ReTurn belt, wipeable, sizes Small through Extra Large, Plastic pouch, 1 item per pouch. Patient transfer aid.
Manufacturer(s)
Handicare USA, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →