Horiba Instruments, Inc dba Horiba Medical recalls 10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400

Recalled February 7, 2018 · FDA recall Z-0493-2018 · Horiba Instruments, Inc

Hazard
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. · FDA Class II
Units
10 bags, 100 units each
Sold at
Nationwide

Description

The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.

Product
10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Manufacturer(s)
Horiba Instruments, Inc dba Horiba Medical