Horiba Instruments, Inc Recalls
17 recalls on record · November 21, 2012 to July 4, 2018 · Medical Devices
Recalls by year
12
13
14
15
18
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
Also filed under
The recall feeds record what looks like the same company under more than one name:
July 4, 2018Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with cert
Hazard: When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
February 7, 2018Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls 15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
Hazard: The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
February 7, 2018Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls 10 ML REAGENT CUP, B1034626, SAP 1221034626 for use with ABX PENTRA 400 / PENTRA C400
Hazard: The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
January 14, 2015Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Reagent Container, Model No. B1037307 (15 mL), B1034626 (10 mL), and B1034634 (4 mL)
Hazard: HORIBA Medical is recalling the ABX PENTRA 400 Reagent Container because they do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. The defect would result in a false high or false low results for HbA1c and false high results for magnes
October 1, 2014Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with associated
Hazard: Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.
March 19, 2014Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls Device Name: ABX PENTRA ML Model/Part #(s): A11A01652 Lot/Serial #(s): N/A Software Version(s): All versions are subj
Hazard: HORIBA Medical initiated this recall of all software versions of ABX PENTRA ML (Model A11A06152) due to a software bug with the potential of transmitting incorrect results to the Laboratory Information System.
March 12, 2014Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical
Hazard: HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers
February 12, 2014Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consist
Hazard: HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to include revised acceptable Target Ranges for Creatinine.No other analyte values are affected.
February 12, 2014Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoring accura
Hazard: Horiba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium.
October 2, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions: intracellula
Hazard: HORIBA Medical is informing all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent.
April 24, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Vers
Hazard: HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.
April 10, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX Micros IM2, Part# M60CSB110EN04 and D00A00202 The IM2 Data Management System consists of the IM2 Data Manager Sof
Hazard: Horiba Medical is recalling ABX Micros IM2 Data Management System because it may display incorrect patient test results under unanticipated operating conditions.
January 16, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2
Hazard: Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software
January 9, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls PENTRA C200, Part Number: A11A01629. The Pentra C200 system is a fully automated chemistry analyzer using colorimetry
Hazard: Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
January 9, 2013Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in
Hazard: Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
November 28, 2012Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16
Hazard: HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.
November 21, 2012Medical DevicesFDA
Horiba Instruments, Inc dba Horiba Medical recalls ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 20
Hazard: This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.