Home / Recalls / Medical Devices
Horiba Instruments, Inc dba Horiba Medical recalls 15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
Recalled February 7, 2018 · FDA recall Z-0494-2018 · Horiba Instruments, Inc
Hazard
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm. · FDA Class II
Units
10 bags of 100 units
Sold at
Nationwide
Description
The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Product
15 ML REAGENT CUP, B1037307, SAP 1221037307 for use with ABX PENTRA 400 / PENTRA C400
Manufacturer(s)
Horiba Instruments, Inc dba Horiba Medical
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →