Ion Beam Applications S.A. recalls Proteus 235

Recalled June 28, 2017 · FDA recall Z-2592-2017 · Ion Beam Applications S.A

Hazard
It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers. · FDA Class II
Units
18 systems ( 10 in the US)
Sold at
Worldwide Distribution - US Distribution to the states of : MA, FL, VA, PA, IL, NJ, WA, TN, LA and TX., and to the countries of: Germany, South Korea, France, Czech Republic, Italy, Poland and Sweden.

Description

It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.

Product
Proteus 235
Manufacturer(s)
Ion Beam Applications S.A.
Made in
Belgium

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov