Home / Recalls / Medical Devices
NRT X-RAY A/S recalls 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combinati
Recalled February 2, 2022 · FDA recall Z-0526-2022 · NRT X-RAY A/S
Hazard
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left. · FDA Class II
Units
72 units
Sold at
USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden
Description
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.
Product
1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combination with Intelli-C devices)
Manufacturer(s)
NRT X-RAY A/S
Made in
Denmark
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →