NRT X-RAY A/S Recalls
4 recalls on record · February 2, 2022 to May 15, 2024 · Medical Devices
Recalls by year
22
23
24
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
May 15, 2024Medical DevicesFDA
NRT X-RAY A/S recalls Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Inte
Hazard: X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT movement is driven at maximum speed and angulated more than 75 deg. from vertical, and 4) the emergency stop or touch guard is activated during the breaking sequence, which could lead to the C-arm hitting the patient or operator.
June 7, 2023Medical DevicesFDA
NRT X-RAY A/S recalls Intelli-C, Left side suspended Tabletop, REF: 03400010
Hazard: There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
June 7, 2023Medical DevicesFDA
NRT X-RAY A/S recalls Intelli-C, Right side suspended Tabletop, REF: 03400000
Hazard: There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
February 2, 2022Medical DevicesFDA
NRT X-RAY A/S recalls 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combinati
Hazard: Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.