Recalled June 7, 2023 · FDA recall Z-1668-2023 · NRT X-RAY A/S

Hazard
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop. · FDA Class II
Units
32 units
Sold at
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.

Description

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

Product
Intelli-C, Right side suspended Tabletop, REF: 03400000
Manufacturer(s)
NRT X-RAY A/S
Made in
Denmark