Philips Electronics North America Corporation recalls Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10

Recalled March 7, 2018 · FDA recall Z-0717-2018 · Philips Electronics North America

Hazard
Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor. · FDA Class II
Units
1486
Sold at
Worldwide Distribution - US Nationwide

Description

Extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them. If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.

Product
Allura Xper FD and and Xper FD ORT - Philips X-ray Systems Product Usage: The Allura Xper FD10 and Alllura Xper FD10/10 is intendd for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and ElectroPhysiology (EP). The Allura Xper FD20, Allura Xper FD20/10, Allura Xper FD20/15 and Allura Xper FD10/10 is intended for: Dedicated vascular and neurovascular imaging applications, including diagnostic and interventional procedures. This includes, e.g. peripheral, cerebral, thoracic and abdominal angiography, as well as PTCAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations and ElectroPhysiology. Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures. Additionally: The Allura Xper FD series is compatible with a hybrid Operating Room (OR Table)
Manufacturer(s)
Philips Electronics North America Corporation