Home / Recalls / Medical Devices
Philips Electronics North America Corporation recalls Allura Xper FD20/15 R8.2 722058 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagn
Recalled February 1, 2017 · FDA recall Z-0925-2017 · Philips Electronics North America
Hazard
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled. · FDA Class II
Units
2
Sold at
US: WA OUS: China, Spain, Korea, Republic of, Germany
Description
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
Product
Allura Xper FD20/15 R8.2 722058 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electro Physiology(EP).
Manufacturer(s)
Philips Electronics North America Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →