Philips Electronics North America Corporation recalls Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnost

Recalled February 1, 2017 · FDA recall Z-0924-2017 · Philips Electronics North America

Hazard
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled. · FDA Class II
Units
2
Sold at
US: WA OUS: China, Spain, Korea, Republic of, Germany

Description

Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.

Product
Allura Xper FD20 R8.2 722028 is intended for: Cardiovascular and vascular X-ray imaging applications, including diagnostic, interventional procedures (such as PTCA, stent placement and atherectomies), pacemaker implantations and Electrophysiology(EP).
Manufacturer(s)
Philips Electronics North America Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov