Philips Electronics North America Corporation recalls Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20 The Allura Xper series i

Recalled December 27, 2017 · FDA recall Z-0255-2018 · Philips Electronics North America

Hazard
When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination. · FDA Class II
Units
157
Sold at
Worldwide Distribution - USA (nationwide) Distribution.

Description

When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.

Product
Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20 The Allura Xper series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures ( such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP) Non-vascular interventions such as drainages, biopsies and vertebroplastics procedures The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.
Manufacturer(s)
Philips Electronics North America Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov