RecallWatchUSA

Home / Recalls / Medical Devices

Philips Electronics North America Corporation recalls Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center

Recalled March 7, 2018 · FDA recall Z-0776-2018 · Philips Electronics North America

Hazard
Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018. · FDA Class II
Units
20357
Sold at
Domestic and Foreign

Description

Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.

Product
Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0
Manufacturer(s)
Philips Electronics North America Corporation