RecallWatchUSA

Home / Recalls / Medical Devices

Philips Electronics North America Corporation recalls Philips Network Firewall (Cisco ASA 5506), Model # 867098

Recalled March 14, 2018 · FDA recall Z-0850-2018 · Philips Electronics North America

Hazard
Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot. · FDA Class II
Units
332 units
Sold at
US Nationwide

Description

Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX. The defect involves the clock signal component within the firewall. This component has a high probability of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable. This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.

Product
Philips Network Firewall (Cisco ASA 5506), Model # 867098
Manufacturer(s)
Philips Electronics North America Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Electronics North America Corporation recalls Philips Network Firewall (Cisco ASA 5506), Model # 867098 | RecallWatchUSA