Home / Recalls / Medical Devices
Philips Electronics North America Corporation recalls Philips SureSigns VS4 Vital Signs Monitor, 863283 863286 The SureSigns VS4 vital signs monitor is for use by health
Recalled August 29, 2018 · FDA recall Z-2894-2018 · Philips Electronics North America
Description
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →