Randox Laboratories Ltd. recalls CK MB Calibrator, Catalogue Number CK2393
Recalled May 25, 2022 · FDA recall Z-1097-2022 · Randox Laboratories
Hazard
Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples. · FDA Class III
Units
1,392
Sold at
Domestic distribution to WV and PR. International distribution to Bangladesh, Bosnia & Herzegovina, Botswana, China, Colombia, Costa Rica, Croatia, Egypt, Fiji, France, Georgia, Germany, Ghana, Greece, Grenada, Hong Kong, India, Iraq, Italy, Lithuania, Malaysia, Maldives, Malta, Mexico, Morocco, Mozambique, Nepal, Nigeria, Pakistan, Philippines, Poland, Puerto Rico, Romania, Russia, Saudi Arabia, South Korea, Switzerland, Thailand, Turkey, UK, Uruguay, Uzbekistan and Vietnam.
Description
Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.
Product
CK MB Calibrator, Catalogue Number CK2393
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland