RecallWatchUSA

Home / Recalls / Medical Devices

Randox Laboratories Ltd. recalls IgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma Catalog Number: IA3832

Recalled December 14, 2022 · FDA recall Z-0362-2023 · Randox Laboratories

Hazard
Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results · FDA Class II
Units
20 kits US in total
Sold at
US Nationwide Distribution: CA, NC, NJ and WV,

Description

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

Product
IgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma Catalog Number: IA3832
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Randox Laboratories Ltd. recalls IgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma Catalog Number: IA3832 | RecallWatchUSA