Randox Laboratories Ltd. recalls Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic us

Recalled August 8, 2018 · FDA recall Z-2575-2018 · Randox Laboratories

Hazard
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring. · FDA Class II
Units
1 kit (US), 602 kits (OUS)
Sold at
Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam

Description

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Product
Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Randox Laboratories Ltd. recalls Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic us | RecallWatchUSA