Home / Recalls / Medical Devices
Randox Laboratories Ltd. recalls Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml
Recalled February 12, 2020 · FDA recall Z-1100-2020 · Randox Laboratories
Hazard
A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86). · FDA Class III
Units
2
Sold at
US: NY and DE OUS: France, Lithuania, Poland, Spain, Switzerland, Turkey, Australia, Hong Kong, Bosnia and Herzegovina, Costa Rica, Ecuador, Gambia, New Zealand and Solomon Islands.
Description
A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).
Product
Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →