Randox Laboratories Ltd. recalls NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Recalled July 28, 2021 · FDA recall Z-2099-2021 · Randox Laboratories

Hazard
Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples · FDA Class II
Units
71 kits US
Sold at
US Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX

Description

Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples

Product
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland