RecallWatchUSA

Home / Recalls / Medical Devices

Randox Laboratories Ltd. recalls RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Recalled December 8, 2021 · FDA recall Z-0328-2022 · Randox Laboratories

Hazard
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range · FDA Class II
Units
15 kits
Sold at
US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.

Description

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Product
RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Randox Laboratories Ltd. recalls RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200 | RecallWatchUSA