Home / Recalls / Medical Devices
Randox Laboratories Ltd. recalls Randox Mg R1 - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catal
Recalled December 16, 2020 · FDA recall Z-0546-2021 · Randox Laboratories
Hazard
Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine · FDA Class II
Units
1 kit
Sold at
US Nationwide distribution in the states of GA, NJ.
Description
Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
Product
Randox Mg R1 - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG3880
Manufacturer(s)
Randox Laboratories Ltd.
Made in
Ireland