RecallWatchUSA

Home / Recalls / Medical Devices

Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a p

Recalled February 5, 2020 · FDA recall Z-0879-2020 · Synthes

Hazard
There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field. · FDA Class II
Units
25 total
Sold at
US Nationwide distribution in the states of AL and ID.

Description

There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.

Product
2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.
Manufacturer(s)
Synthes, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a p | RecallWatchUSA