Synthes Recalls

43 recalls on record · February 26, 2014 to August 5, 2020 · Medical Devices

Recalls by year

14
15
18
20

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

August 5, 2020Medical DevicesFDA
Synthes, Inc. recalls MatrixNEURO Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton
Hazard: One lot of MatrixNEURO 5mm screws was packaged in a body clip etched "4", which indicated a 4mm screw. The outer package labeling correctly identifies a 5mm screw, but the body clop incorrectly identifies a shorter length for the screw.
February 5, 2020Medical DevicesFDA
Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 24MM STERILE, Part Number 04.210.124TS - Product Usage:Sterile Tube Packaging is a p
Hazard: There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
February 5, 2020Medical DevicesFDA
Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile Tube Packaging is a p
Hazard: There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
February 5, 2020Medical DevicesFDA
Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 16MM STERILE, Part Number 04.210.116TS - Product Usage:Sterile Tube Packaging is a p
Hazard: There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
February 5, 2020Medical DevicesFDA
Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 12MM STERILE, Part Number 04.210.112TS - Product Usage:Sterile Tube Packaging is a p
Hazard: There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
February 5, 2020Medical DevicesFDA
Synthes, Inc. recalls 2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile Tube Packaging is a p
Hazard: There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.
May 2, 2018Medical DevicesFDA
Synthes, Inc. recalls Stem Extractor f/Guide Bar, Part number 03.401.072. Instrumentation for extraction and revision of Epoca shoulder impla
Hazard: There is a potential for the connection screw of the stem extractor to break.
May 2, 2018Medical DevicesFDA
Synthes, Inc. recalls Flex Arm Product Usage: The Synthes Flex Arm is a component of the Synthes Minimally Invasive Support System, which
Hazard: Quick connect feature of instruments may be unable to connect to the male features of Flex Arm Adaptors, Insight Retractors, or other mating parts.
February 28, 2018Medical DevicesFDA
Synthes, Inc. recalls LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation o
Hazard: May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
July 22, 2015Medical DevicesFDA
Synthes, Inc. recalls Reamer / Irrigator / Aspirator (RIA): RIA Drive Shafts, RIA Tube Assemblies, sterile, RIA Reamer Head, sterile (12.0mm-
Hazard: It was reported that the RIA Drive Shaft, Tube Assembly, and Reamer Head have the potential to break when incorrectly assembled or used improperly. Precautionary statements are being added to the Reamer / Irrigator / Aspirator (RIA) Surgical Technique Guide.
July 15, 2015Medical DevicesFDA
Synthes, Inc. recalls TITANIUM END CAP WITH T40 STARDRIVE RECESS, 0MM EXTENSION FOR TITANIUM TIBIAL NAILS-EX; Synthes Tibial Nail System EX is
Hazard: A specific part and lot number was packaged and shipped prior to the completion of a required internal inspection.
July 8, 2015Medical DevicesFDA
Synthes, Inc. recalls DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm, 75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Exp
Hazard: certain lots of the Titanium Recon Screws (6.5mm ) were found to contain the incorrect insert GP0760-A (Synthes Titanium Buttress Locking Plate System), instead of the correct insert GP0706-E (Synthes Titanium Intramedullary Nails and Components).
June 24, 2015Medical DevicesFDA
Synthes, Inc. recalls 3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated for fixation
Hazard: Certain lots of the 3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws were found to have been labeled incorrectly.
June 24, 2015Medical DevicesFDA
Synthes, Inc. recalls Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is placed throug
Hazard: Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage.
June 10, 2015Medical DevicesFDA
Synthes, Inc. recalls 3.2 mm Cannulated Drill Bit/ QC 170 mm 2.7 Cannulated Drill Bit/ QC 160 mm 2.0 mm Cannulated Drill Bit / QC 150 mm t
Hazard: affected parts and lots of the Cannulated Drill Bits have the potential to break during use
May 20, 2015Medical DevicesFDA
Synthes, Inc. recalls TFN-ADVANCED Proximal Femoral Nailing System (TFNA); for treatment of fractures in orthopedics and trauma.
Hazard: TFNA nails from certain lots were assembled with a locking mechanism too close to the top of the nail. Potential patient impact: surgical delay.
April 8, 2015Medical DevicesFDA
Synthes, Inc. recalls Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part
Hazard: The guide sleeve has the potential to jam in the Aiming Arms, which may delay disassembly and removal of instruments.
April 1, 2015Medical DevicesFDA
Synthes, Inc. recalls Synthes Electric Pen Drive (EPD) System, The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire r
Hazard: In May 2013, the original recall was initiated due to complaints the Hand Switch for the Electric Pen Drive was not labeled correctly, and the on and lock labels were interchanged.
March 25, 2015Medical DevicesFDA
Synthes, Inc. recalls Thumb Screw for the Synthes Recon Locking Aiming Arm for Lateral Entry Femoral Nails-EX (part number 03.010.048) and the
Hazard: For certain lots, the incorrect raw material of annealed 17-4PH Stainless Steel was used instead of the correct material of heat-treated 304 Stainless Steel. This may lead to stress-corrosion cracking, which can lead to surgical delay or unintended debris entering the surgical site.
March 18, 2015Medical DevicesFDA
Synthes, Inc. recalls Synthes Inserter for Titanium Elastic Nails; The Inserter for Titanium Elastic Nails (TEN) is intended for fixation of d
Hazard: The Inserter has the potential for mechanical failures such as breakage of the main shaft of the Inserter for the TEN, malfunction of the chuck of the Inserter for the TEN in the form of the chuck getting stuck/jammed, breakage of the cross bar of the Inserter for the TEN.
February 18, 2015Medical DevicesFDA
Synthes, Inc. recalls Hemostatic Bone Putty, for use as a water-soluble implant material and for use in the control of bleeding from bone sur
Hazard: Revision A of the Hemostatic Bone Putty Brochure (J10847A) was sent to sales consultants instead of Revision B (J10847B), which corrected an error in the indication statement.
February 4, 2015Medical DevicesFDA
Synthes, Inc. recalls 2.0 Mandible Locking Plate (MLP) Bender / Cutter (part number 329.143) Intended to be used for bending and/or cutting
Hazard: Certain lots have the potential for tip breakage while bending 2.0 mm thick plates. If device is used and breaks while in the Operating Room, surgical delay may occur while alternate bending or cutting instruments are located.
February 4, 2015Medical DevicesFDA
Synthes, Inc. recalls The Synthes Mandible External Fixator; all lots of part nos.: 04.305.003 04.305.004 04.305.005 04.305.010 04.305
Hazard: Under new testing protocols for MR Environment safety and compatibility, metal devices are no longer designated MR Safe. Some components in the CMF Mandible External Fixator Systems were labeled and/or etched with information indicating they are MR Safe. They can no longer be labeled MR Safe.
January 21, 2015Medical DevicesFDA
Synthes, Inc. recalls Malleable C- Retractor, Part Number 397.232, used to provide retraction of the soft tissues during use of trocar instrum
Hazard: Potential failure and/or corrosion of the internal spring of the retractor, which may have been manufactured from an incorrect material. Surgical delay may occur. Spring particles could transfer from the instrument to the patient, and may lead to an adverse tissue reaction and/or infection.
November 26, 2014Medical DevicesFDA
Synthes, Inc. recalls 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM - Sterile: intended to treat stable and unstable
Hazard: The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.
October 29, 2014Medical DevicesFDA
Synthes, Inc. recalls DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715
Hazard: Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.
October 1, 2014Medical DevicesFDA
Synthes, Inc. recalls DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate
Hazard: DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems. The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i
August 27, 2014Medical DevicesFDA
Synthes, Inc. recalls Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set
Hazard: A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.
August 27, 2014Medical DevicesFDA
Synthes, Inc. recalls DePuy Synthes Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies)
Hazard: DePuy Synthes is initiating a recall of certain lots of the Craniomaxillofacial Distraction System (AB Distractor Bodies and BC Distractor Bodies) because they may reverse post-operatively.
July 30, 2014Medical DevicesFDA
Synthes, Inc. recalls Synthes Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System). Orthopedic surgical instrum
Hazard: The Guide Blocks for the 2 Column Plate 6 Hole Head (Volar Distal Radius Plate System) were found to be labeled incorrectly. Part no. 03.111.600 (Right) was found in the package labeled part no. 03.111.601 (Left). The laser etching denoting the orientation on the part is correct.