Synthes, Inc. recalls LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation o

Recalled February 28, 2018 · FDA recall Z-0669-2018 · Synthes

Hazard
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening. · FDA Class II
Units
567 units
Sold at
Distribution to US, Canada, and Germany.

Description

May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.

Product
LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Manufacturer(s)
Synthes, Inc.