RecallWatchUSA

Home / Recalls / Medical Devices

Thoratec Corp. recalls Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System

Recalled August 28, 2019 · FDA recall Z-2348-2019 · Thoratec

Hazard
Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure. · FDA Class II
Units
7200 left ventricular assist systems
Sold at
Worldwide distribution. US nationwide, Australia, Austria, Brazil, Canada, Cayman Islands, Chile, Colombia, Egypt, France, Germany, Greece, Singapore, Iran, Italy, Kuwait, Lebanon, Romania, Switzerland, Thailand, United Kingdom.

Description

Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.

Product
Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System
Manufacturer(s)
Thoratec Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Thoratec Corp. recalls Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System | RecallWatchUSA