June 24, 2026Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HM 11V Li-ion Backup Battery, ABT Model/Catalog Number: 106128 Software Version:
Hazard: Due to 11 volt Backup Battery failures.
June 24, 2026Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N
Hazard: Due to 11 volt Backup Battery failures.
June 24, 2026Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US So
Hazard: Due to 11 volt Backup Battery failures.
January 21, 2026Medical DevicesFDA
Thoratec LLC recalls CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre
Hazard: Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
November 26, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿
Hazard: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
November 26, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Ki
Hazard: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
November 26, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Co
Hazard: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
November 26, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, He
Hazard: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
August 20, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: Mobile Power Unit Model/Catalog Number: 107754 Software Version: N/A Product Des
Hazard: Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
August 20, 2025Medical DevicesFDA
Thoratec LLC recalls Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Software Version: N
Hazard: Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.
April 23, 2025Medical DevicesFDA
Thoratec LLC recalls HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with Hear
Hazard: Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.
August 21, 2024Medical DevicesFDA
Thoratec LLC recalls HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant K
Hazard: Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop
June 19, 2024Medical DevicesFDA
Thoratec Corp. recalls Thoratec HeartMate System Monitor, REF: 1286, L1286, 1286A, L1286A, 1286A-US, L1286A-US, 1286C, L1286C, L1386, 1286A-CAN
Hazard: System monitor screen may display atypical behavior: Overlapping screens/buttons, Frozen screen, Distorted text, blanks, or zeroes in place of values, Unresponsive buttons where the user is unable to initiate a command, which may lead to anxiety /inconvenience, hemodynamic compromise, extended surgical time, and Pump Stop command may be selected
May 8, 2024Medical DevicesFDA
Thoratec Corp. recalls Thoratec HeartMate 3, LVAS Implant Kit, REF:106524INT, Sterile EO, RxOnly
Hazard: Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
May 8, 2024Medical DevicesFDA
Thoratec Corp. recalls Thoratec HeartMate 3, LVAS Implant Kit, REF:106524US, Sterile EO, RxOnly
Hazard: Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
March 27, 2024Medical DevicesFDA
Thoratec Corp. recalls HeartMate 3 Left Ventricular Assist System (LVAS): 1. HeartMate 3 Sealed Outflow Graft with Bend Relief, Investigationa
Hazard: Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
March 27, 2024Medical DevicesFDA
Thoratec Corp. recalls HeartMate II Left Ventricular Assist System (LVAS): 1. HeartMate II Sealed Outflow Graft with Bend Relief REF 103393 2
Hazard: Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
February 14, 2024Medical DevicesFDA
Thoratec Corp. recalls 1. HeartMate Touch Communication System, REF HMT1100 2. HeartMate Touch Communication System, Rental REF HMT1100-R 3.
Hazard: Due to software and controller systems that results in unexpected pump start or pump stop.
October 11, 2023Medical DevicesFDA
Thoratec Corp. recalls Heartmate 3 LVAS Implant Kit, REF: 106524US, 106524INT; Heartmate XVE & HeartMate II Apical Coring Knife, REF: 1050
Hazard: Inability to start and/or complete the coring procedure due to inadequate sharpness of the Apical Coring Knife, provided with Left Ventricular Assist System, could result in an extended procedure or surgical time while the backup coring knife is acquired.
December 8, 2021Medical DevicesFDA
Thoratec Corp. recalls Abbott HeartMate Touch Communication System, REF: HMT1100 (US), HMT1100-R (Rental US), HMT1150 (EU)
Hazard: If LVAS communication system is trying to establish Bluetooth connection and another Bluetooth-enabled device is nearby and advertising for Bluetooth connection, the Bluetooth connectivity interference may cause the system App to close or fail to open. Once Bluetooth adaptor is connected, barring disconnection, all functionality is unaffected; the device will continue to operate as expected.
January 8, 2020Medical DevicesFDA
Thoratec Corp. recalls HeartMate 3 Mobile Power Unit, REF 107754, UDI or GTIN: 00813024010883, and 107758, UDI or GTIN: 00813024010890, when co
Hazard: Excessive static electricity can potentially cause unrecoverable power loss and damage to the mobile power unit that can be used to power the left ventricular assist system, which could lead to no or reduced blood flow from the left ventricular assist system.
August 28, 2019Medical DevicesFDA
Thoratec Corp. recalls Thoratec Heartmate 3 Sealed Outflow Graft with Bend Relief, Sterile, REF 105581US/105581, RX only, Part of Left Ventricu
Hazard: Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
August 28, 2019Medical DevicesFDA
Thoratec Corp. recalls Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System
Hazard: Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.
May 1, 2019Medical DevicesFDA
Thoratec Corp. recalls Thoratec HeartMate 3 Coring Tool, Sterile EO, Rx Only, REF 10005872, an optional accessory of Thoratec HeartMate 3 Left
Hazard: Small black plastic particle was observed in the left ventricle during implant procedure.
May 31, 2017Medical DevicesFDA
Thoratec Corporation recalls HeartMate¿ II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to pr
Hazard: St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
May 31, 2017Medical DevicesFDA
Thoratec Corporation recalls HeartMate¿ II LVAS with Pocket Controller; 106015 - HMII LVAD SEALED KIT-PCTRL, NA UDI: 00813024011224 Intended to
Hazard: St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
May 31, 2017Medical DevicesFDA
Thoratec Corporation recalls HeartMate¿ II LVAS with Pocket Controller; 107801 - HMII PUMP & POCKET CTRL ONLY UDI:00813024011286 Intended to pr
Hazard: St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
April 2, 2014Medical DevicesFDA
Thoratec Corporation recalls HeartMate II LVAD Pump and Pock Controller Kit; Product Usage: HeartMate II is a mechanical device that circulates
Hazard: Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
April 2, 2014Medical DevicesFDA
Thoratec Corporation recalls Heart Mate II Pocket Controllers removed from packaging Product Usage: HeartMate II is a mechanical device that ci
Hazard: Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
April 2, 2014Medical DevicesFDA
Thoratec Corporation recalls HeartMate II System Controller Product Usage: HeartMate II is a mechanical device that circulates blood throughout t
Hazard: Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.