RecallWatchUSA

Home / Recalls / Medical Devices

Thoratec LLC recalls HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant K

Recalled August 21, 2024 · FDA recall Z-2583-2024 · Thoratec

Hazard
Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop · FDA Class II
Units
694
Sold at
US: MN, PA, FL, IL, NY, DE, AR, NJ, NC, GA, WI, AZ, TN, MO, TX, MA, OH, VA, NE, CA, OK, CO, DC, IN, MI, CT, UT, IA, MD, OR, NM, ME, SC, KY, LA, WA, NV, HI, AL, MS, KS, WV. OUS: Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan

Description

Left ventricular assist system controller UI membrane/screen may lift along the edge of the controller housing, and if controller is exposed to fluid, ingress may occur, which may damage printed circuit board assemblies, which could result in unexpected and false positive alarms, loss of visual alarms, loss of user interface(use of buttons, loss of visual alarms,etc.), loss of power, and pump stop

Product
HeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS) Implant Kits or distributed separately: HeartMate 3 LVAS Implant Kit, US, Model: 106524US; HeartMate 3 LVAS Implant Kit, OUS, Model: 106524INT; HeartMate 3 System Controller, US, Model: 106531US; HeartMate 3 System Controller, OUS, Model: 106531INT; HeartMate 3 System Controller Low Flow 2.0, Global, Model: 106531LF2
Manufacturer(s)
Thoratec LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov