Home / Recalls / Medical Devices
Tosoh Bioscience Inc recalls AIA-PACK TSH 3rd Generation; Part Number: 020294 Assay, Thyroid
Recalled June 13, 2018 · FDA recall Z-2077-2018 · Tosoh Bioscience
Hazard
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay. · FDA Class II
Units
5293
Sold at
Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.
Description
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.
Product
AIA-PACK TSH 3rd Generation; Part Number: 020294 Assay, Thyroid
Manufacturer(s)
Tosoh Bioscience Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →