RecallWatchUSA

Home / Recalls / Medical Devices

Tosoh Bioscience Inc recalls ST-AIA PACK PAP; Part Number: 025262 Assay, Prostate

Recalled June 13, 2018 · FDA recall Z-2068-2018 · Tosoh Bioscience

Hazard
Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay. · FDA Class II
Units
1083
Sold at
Worldwide Distribution: US (Nationwide) and countries of; Canada, Mexico, Columbia, Chile, Uruguay, Nicaragua, Argentina, Peru, Chile, Honduras. Equador, Caymen Islands, Panama, and Costa Rica.

Description

Asfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

Product
ST-AIA PACK PAP; Part Number: 025262 Assay, Prostate
Manufacturer(s)
Tosoh Bioscience Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov