RecallWatchUSA

Home / Recalls / Medical Devices

Vyaire Medical recalls Infant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number

Recalled January 31, 2018 · FDA recall Z-0344-2018 · Vyaire Medical

Hazard
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient. · FDA Class I
Units
16,670 units in total
Sold at
US Nationwide Distribution

Description

Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

Product
Infant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Manufacturer(s)
Vyaire Medical

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Vyaire Medical recalls Infant AirLife Dual-,limb, dual-heat, high-flow circuit (Model AH265) with AirLife Humidification Chamber, Model Number | RecallWatchUSA