Home / Recalls / Medical Devices
Zeiss, Carl Inc recalls Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Recalled April 15, 2020 · FDA recall Z-1629-2020 · Zeiss, Carl
Hazard
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system · FDA Class II
Units
US: 600; Global: 1600
Sold at
Worldwide Distribution - US Nationwide
Description
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
Product
Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Manufacturer(s)
Zeiss, Carl Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →