RecallWatchUSA

Home / Brands / Zeiss, Carl

Zeiss, Carl Recalls

6 recalls on record · July 24, 2013 to August 3, 2022 · Medical Devices

Recalls by year

13
14
18
20
22

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

August 3, 2022Medical DevicesFDA
Zeiss, Carl Inc recalls ZEISS Axio Observer 5 microscope, Product material no. 431006-9901-000
Hazard: An incorrect Unique Device Identification (UDI) label was installed on some ZEISS Axio Observer 5 microscopes.
April 15, 2020Medical DevicesFDA
Zeiss, Carl Inc recalls Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Hazard: When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
July 25, 2018Medical DevicesFDA
Zeiss, Carl Inc recalls Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Hazard: Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
June 20, 2018Medical DevicesFDA
Zeiss, Carl Inc recalls Cell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning
Hazard: In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
January 29, 2014Medical DevicesFDA
Zeiss, Carl Inc recalls Compensator Slider Lambda 6x20 component (part number 473704-0000-000) utilized on all upright ZEISS microscopes (Axio I
Hazard: Carl Zeiss Microscopy, LLC has recalled the Compensator Slider Lamda 6x20 component utilized on Standard Pathology Microscopes due to misalignment of the lambda plate.
July 24, 2013Medical DevicesFDA
Zeiss, Carl Inc recalls Polarizer D Fixed w/Rotating Red 1 lambda plate. Utilized in upright Zeiss microscopes to aid in diagnosis by means o
Hazard: Carl Zeiss Microscopy, LLC is recalling the Polarizer D component used in All upright ZEISS microscopes (Axio Imager, Axio Scope.A1, Axio Lab.A1) due to Misaligned Full-Wave (Lambda) Plate.
Zeiss, Carl Recalls | RecallWatchUSA