Home / Recalls / Medical Devices
Zeiss, Carl Inc recalls Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Recalled July 25, 2018 · FDA recall Z-2441-2018 · Zeiss, Carl
Hazard
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation. · FDA Class II
Units
8
Sold at
US Nationwide Distribution
Description
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Product
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Manufacturer(s)
Zeiss, Carl Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →