RecallWatchUSA

Home / Recalls / Medical Devices

Zeiss, Carl Inc recalls Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Recalled July 25, 2018 · FDA recall Z-2441-2018 · Zeiss, Carl

Hazard
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation. · FDA Class II
Units
8
Sold at
US Nationwide Distribution

Description

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

Product
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Manufacturer(s)
Zeiss, Carl Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Zeiss, Carl Inc recalls Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope | RecallWatchUSA