127,502 recalls from the U.S. Consumer Product Safety Commission, newest first.

February 12, 2025Medical DevicesFDA
Ethicon, LLC recalls PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general so
Hazard: Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
February 12, 2025Medical DevicesFDA
Hologic, Inc. recalls The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 7173
Hazard: Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN¿ for the TLiIQ¿ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 5 M12 System Code: (1) 722227 (2) 722231
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 M12 System Code: (1) 722063 (2) 722221
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXperFD20/15 System Code: (1) 722058
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/20 System Code: (1) 722038
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXperFD20/10 System Code: (1) 722029
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXper FD20 OR Table System Code: (1) 722023 (2) 722035
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10C System Code: (1) 722001
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls AlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura CV20 System Code: 722031
Hazard: Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
February 12, 2025Medical DevicesFDA
Oxoid Limited recalls Lev5, REF CT1587B, 50 Susceptibility discs Levofloxacin 5 ug, IVD, CE 2797
Hazard: Mislabeling; Levofloxacin Antimicrobial susceptibility disc packaging may contain Norfloxacin Antimicrobial Susceptibility discs.
February 12, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimul
Hazard: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
February 12, 2025Medical DevicesFDA
LivaNova USA, Inc. recalls SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimula
Hazard: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
February 12, 2025Medical DevicesFDA
ROi CPS LLC recalls regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to
Hazard: There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
February 12, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls Optima XR646 HD. X-Ray imaging system.
Hazard: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
February 12, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls Discovery XR656 HD. X-Ray imaging system.
Hazard: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
February 12, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls Discovery XR656HD. X-Ray imaging system.
Hazard: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
February 12, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
Hazard: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
February 12, 2025Medical DevicesFDA
Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care enviro
Hazard: Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.