April 19, 2017Medical DevicesFDA
ConMed Corporation recalls HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION,
Hazard: For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
April 19, 2017Medical DevicesFDA
ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurg
Hazard: For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
April 19, 2017Medical DevicesFDA
ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurg
Hazard: For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
April 19, 2017Medical DevicesFDA
ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurg
Hazard: For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
April 19, 2017Medical DevicesFDA
ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurg
Hazard: For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Trumpet Handpiece only, Catalog Number CD8450 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in g
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Trumpet Handpiece with 5mm x 32cm Length Probe , Catalog Number CD8400 The CORE Trumpet (CD81XX, CD83XX, and CD84
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Suction Irrigation Handpiece without Probe for Single Solution or Dual Bags, Catalog Number CD8302 The CORE Trump
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Suction Irrigation Handpiece without Probe for Single Solution Bags, Catalog Number CD8300 The CORE Trumpet (CD81
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Suction Irrigation Handpiece Y-Tubing set, with 5mm x 32cm Length Probe for Single Solution or Dual Bags, Catalog
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Suction Irrigation Handpiece with 5mm x 32cm Length Probe for Single Solution or Dual Bags, Catalog Number CD8190
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 26, 2016Medical DevicesFDA
ConMed Corporation recalls CORE¿ Suction Irrigation Handpiece with 5mm x 32cm Length Probe for Single Solution Bags, Catalog Number CD8185 The COR
Hazard: The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.
October 19, 2016Medical DevicesFDA
ConMed Corporation recalls CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The H
Hazard: CONMED received complaints of damage to the CS-023 electrodes shaft insulation which could result in burn marks on the shaft. If the shaft insulation is damaged in any way, it could increase the risk of patient burn.
August 31, 2016Medical DevicesFDA
Conmed Corporation recalls Pediatric Dual Dispersive Electrodes (for patients 5-15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birt
Hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
August 31, 2016Medical DevicesFDA
Conmed Corporation recalls Adult Dual Dispersive Electrodes (for patients >15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birtcher
Hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
August 31, 2016Medical DevicesFDA
Conmed Corporation recalls Pediatric ThermoGard¿ Dual Dispersive Electrodes (for patients 5-15 kg.), 10' (3.05m) Cable, Catalog Number 51-7710.
Hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
August 31, 2016Medical DevicesFDA
Conmed Corporation recalls Adult ThermoGard¿ Dual Dispersive Electrodes (for patients >15 kg.), No Cable, Catalog Number 51-7410. Intended to be
Hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
August 31, 2016Medical DevicesFDA
Conmed Corporation recalls Adult ThermoGard¿ Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number 51-7310. Inten
Hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
June 8, 2016Medical DevicesFDA
ConMed Corporation recalls Hip Preservation System Signature Series PreBent Burs under the following labels: 1) PreBent Spherical Bur, 5.5mm x 19cm
Hazard: Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.
December 9, 2015Medical DevicesFDA
ConMed Corporation recalls Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutt
Hazard: Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.
November 11, 2015Medical DevicesFDA
ConMed Corporation recalls Arthroscopic Energy 50¿ Probe with Suction, Catalog number: AES-50S Peoduct Usage: For resection, ablation, and coag
Hazard: Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls CONMED ADULT R2 Multifunction Electrodes, REF/Catalog Number 3115-1751, Direct Connect to Philips Heartstart, Rx ONLY --
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls CONMED PEDIATRIC R2 Multifunction Electrodes, REF/Catalog Number 3115-1750, Direct Connect to Philips Heartstart, Rx ONL
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls MINI PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2602H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTAR
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2603H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELE
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls ADULT Radiotransparent Electrode, REF/Catalog Number 2516H-PC, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELEC
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls ADULT Radiotransparent Electrode, REF/Catalog Number 2516H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRO
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-PC, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELEC
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls ADULT Radiotransparent Electrode, REF/Catalog Number 2001H-C, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECT
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
December 10, 2014Medical DevicesFDA
ConMed Corporation recalls ADULT Radiotransparent Electrode, REF/Catalog Number 2001H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRO
Hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.