April 12, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AS
Hazard: Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.
April 5, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500 Intelligent Lab System and Dimension Vista 1500 Intelligent Lab System, analyzer, chemistry (photome
Hazard: Siemens Healthcare Diagnostics has confirmed that in isolated cases when ECREA is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result.
March 29, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur XPT System Software Versions V1.0.1, V1.0.2, V1.0.3, V1.1 and V1.2. The ADVIA Centaur XPT System is a stan
Hazard: Multiple software issues, which may affect the operation and workflow of the system. Potential for an apparent delay to testing when these issues occur. The software issues affecting the analyzer may impact all analytes available on the ADVIA Centaur XPT System test menu.
March 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls IMMULITE 2000 Progesterone; Catalog Number: L2KPW2 L2KPW6; Siemens Material Number: 10381181, 10381170 Product Usag
Hazard: DHEA-S causes falsely elevated progesterone results.
March 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in
Hazard: DHEA-S causes falsely elevated progesterone results.
March 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls Dimension Vista LOCI Progesterone; Catalog Number: K6464; Siemens Material Number: 10461743 Product Usage: For in vi
Hazard: DHEA-S causes falsely elevated progesterone results.
March 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur Progesterone; Catalog Number: 10310305 10315522 10333111 Product Usage: The PROG method is an in vitro
Hazard: DHEA-S causes falsely elevated progesterone results.
March 8, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904
Hazard: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.
March 8, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904
Hazard: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.
March 8, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls Siemens ADVIA Centaur Systems Calibrator A Kits, Cat No. 04800646, SMN 10285903 (2 pack)Cat No. 04800735, SMN 10285904
Hazard: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur FT4 when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT Systems. In addition, Siemens Healthcare Diagnostics confirmed the potential for calibration failures due to above limit calibrator RLU %CVs when using Calibrator A kit lots ending in 90 with the FT4 assay. The performance of the FT4 assay when used with Calibrator A kit lots ending in 90 on the ADVIA Centaur CP System is not affected. Customers may continue to use Calibrator A kit lots ending in 90 for the ADVIA Centaur Systems FT3, T3, T4 and TUp assays.
March 1, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls N Latex HCY REAGENT, REF/Catalog OPAX03 /SMN 10445973 , IVD Product Usage: In-vitro diagnostic reagents for the qu
Hazard: The firm confirmed a reduced once-opened and on-board stability for N Latex Homocysteine (HCY) OPAX03 lot 802907632 that may result in erroneously reduced or elevated homocysteine values. The stability issue may lead to higher than expected lot-to-lot variation and an impaired product performance.
February 22, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ cli
Hazard: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
February 22, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI
Hazard: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
February 22, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solu
Hazard: There is a remote possibility CentraLink may download an order to the ADVIA Automation System without specifying the sample type. This can occur when an order is received from the LIS without a sample type, requiring that the sample type be set in CentraLink based on the sample type of the test in the order.
February 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the
Hazard: Siemens Healthcare Diagnostics has determined that Dimension Vista AMM (K3119) Flex reagent cartridge flex lots 16187BE, 16225BB, 16265AB do not meet the 60-day calibration interval claim due to reagent instability and results may show an Abnormal Assay.
February 15, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension¿ Ammonia Flex¿ reagent cartridge / AMM Dimension: The AMM method is an in vitro diagnostic test for the quan
Hazard: Siemens Healthcare Diagnostics has determined that Dimension AMM (DF119) flex reagent cartridge lots EB7180, FB7152, BA7194, EA7223, BA7250 do not meet the 60-day calibration interval claim due to reagent instability and results may show an Abnormal Assay.
February 8, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LK3PS1, SMN 10706293 (US Kit Catalog) Produc
Hazard: Negative bias results with the IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit was confirmed for two lots (D0119 and D0120). The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
February 8, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems, Third Generation PSA, Catalog Number/REF LKUP1, SMN 10380956 (OUS Kit Catalog) Produc
Hazard: Negative bias with IMMULITE¿ /IMMULITE¿ 1000 Third Generation PSA (prostate-specific antigen) kit. The investigation by Siemens has confirmed that patient samples exhibit an average negative bias of approximately -22% with a range of -16% to -27% for samples recovering from 0.08 to 16.2 ng/mL.
January 18, 2017Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls SIEMENS IMMULITE 2000/2000 Xpi Systems Intact PTH (Parathyroid Hormone) Test Code iPTH, REF/Catalog Number L2KPP2, S
Hazard: Siemens is conducting this recall for the IMMULITE¿ 2000/IMMULITE¿ 2000 XPi Intact PTH (Intact Parathyroid Hormone) (iPTH) assay kit lot 320. Siemens has confirmed that IMMULITE¿ 2000/IMMULITE¿ 2000 XPi Intact PTH kit lot 320 can exhibit an average negative bias of up to -39% at iPTH concentrations <20 pg/mL with serum and EDTA patient samples vs. a reference kit lot.
December 28, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000/IMMULITE 2000 XPi Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number L2KPTS2, SMN 10706281
Hazard: Siemens Healthcare Diagnostics confirmed that the IMMULITE 2000/ IMMULITE 2000 XPi Systems PSA assays are not meeting the current High-dose Hook Effect expectation in the Instructions for Use.
December 28, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number
Hazard: Siemens Healthcare Diagnostics has confirmed that the IMMULITE/ IMMULITE 1000 Systems PSA assays are not meeting the current High-dose Hook Effect expectation in the Instructions for Use.
December 14, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls ADVIA 560 Hematology Systems, Siemens Material Number (SMN) 11170842, IVD.
Hazard: Siemens is investigating an issue which may cause an incorrect result to be reported. Siemens received two reports of multiple discordant records for the same Sample ID in the ADVIA 560 Hematology System database that occurred during the installation of the systems. The database should only contain one record of a Sample ID number for any given time and date. If there are multiple records for the same Sample ID, it is possible that multiple results may be manually or automatically sent to the Laboratory Information System (LIS), printed or displayed on the results report screen.
November 30, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista Chemistry 1 Calibrator The Chemistry 1 Calibrator (KC110B) is an in vitro diagnostic product for the ca
Hazard: Siemens determined that discrepant low calcium recovery of QC and patient samples can occur with Dimension Vista¿ CHEM 1 CAL lots 5GM081 and 5GM082.
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur CA 19-9 Assay (250 test kit)-for use,with the ADVIA Centaur and ADVIA Centaur XP, ADVIA Centaur XPT and A
Hazard: CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc recalls ADVIA Centaur CA 19-9 Assay (50 test kit)-for use, with the ADVIA Centaur and ADVIA Centaur XP, ADVIA Centaur XPT and A
Hazard: CA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls FlexLab Automation Modules
Hazard: FDA Class II
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Aptio Automation Modules
Hazard: Possible overheating of power cord plugs used to connect to FlexLab and Aptio Automation systems.
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the followin
Hazard: Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension IRON Flex reagent cartridges
Hazard: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).
November 23, 2016Medical DevicesFDA
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista IRON Flex reagent cartridges
Hazard: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).