Home / Brands / Smiths Medical ASD
Smiths Medical ASD Recalls
293 recalls on record · July 18, 2012 to May 14, 2025 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
19
20
21
22
23
24
25
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) PAED. TRACHEOSTOMY TUBE 2.5MM UNCUFFED , Product Code/List N
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) 3.0MM FLEXTEND NEO STRAIGHT TRACH TUBE , Product Code/List
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: 2.5MM UNCUFFED NEONATAL FLEXTEND PLUS TRACHEOSTOMY TUBE W/NEO S
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE 2.5FLEXTEND PLUS , Product Code/L
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACH TUBE , Product Code/List Number/Item Code 60N
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) NEONATAL TRACHEOSTOMY TUBE , Product Code/List Number/Item C
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) AIRE-CUF TRACHY TUBE 6.0MM MID-RANGE , Product Code/List Num
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
September 4, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORTEX Bivona Tracheostomy Silicone Tube, labeled as: a) TTS TRACHEOSTOMY TUBE 7.0MM TIGHT TO SHAFT CUFF, Product Cod
Hazard: Smiths Medical has identified that the securement flange of specific lots of the Bivona Neonatal/Pediatric and Adult Tracheostomy products may tear because of a manufacturing defect.
August 14, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD-Solis VIP ambulatory Infusion Pumps, Model 2120 (21-2120, 21-2125, and 21-2127) Indicated for the following u
Hazard: The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
August 14, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD-Solis VIP ambulatory Infusion Pumps, Model 2110. Indicated for the following uses: Intra-arterial, subcutaneous,
Hazard: The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
August 14, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD-Solis Ambulatory Infusion Pumps, Model 2100. Intended for the following uses: Indicated for the following uses: I
Hazard: CADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air Detection, 5. Error Codes Not Displayed at Power Up, 6. Audible Alarm, 7. Low Sensitivity Air in Line Detection Threshold, 8. PharmGuard Server Password. Smiths Medical corrected many of the issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent CADD software (v4.3 - 2023) installed on your pumps.
April 24, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716
Hazard: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
April 24, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 301
Hazard: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
April 24, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-5
Hazard: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
April 24, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0
Hazard: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
March 27, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls smiths medical Pneupac paraPAC plus 300 ventilator kit without PEEP and CPAP, REF P300NUS
Hazard: When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous positive gas flow when in the cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.
March 27, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls smiths medical Pneupac paraPAC plus 310 ventilator kit with PEEP and CPAP, REF P310NUS
Hazard: When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous positive gas flow when in the cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.
February 7, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls smiths medical Medfusion Model 4000 syringe pump
Hazard: Medfusion Model 4000 syringe infusion pumps, with the following software versions, may have the following issues: 1. Delivery During Motor Not Running High Priority Alarm (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 2. Re-administered Loading Dose (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 3. Incorrect Critical Data Failure Alarm, 4. Interruption of Bolus or Loading Dose Delivery (v1.0.0, v1.1.0, v1.1.1, v1.1.2), 5. Incorrect Total Bolus/Loading Dose Displayed (v1.1.0, v1.1.1, v1.1.2), 6. Volume Limit Before Bolus/Loading Dose Complete (v1.1.2), 7. Drug Library Lower Limit Displayed Incorrectly (v1.6.0, v1.6.1), 8. Depleted Battery Alarm (All versions previous to v1.6.5), 9. Loss of Wireless Connectivity (v1.5.0, v1.5.1, v1.6.0, v1.6.1, v1.6.4), 10. PharmGuard Server Password (v2.3, v2.4, v2.5). Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software installed on your pumps.
February 7, 2024Medical DevicesFDA
Smiths Medical ASD Inc. recalls smiths medical Medfusion Model 3500 Syringe pump
Hazard: Medfusion syringe pumps, with software versions before v6.0.0, may have the following issues:1. Delivery During Motor Not Running High Priority Alarm, 2. Infusion Restarted with Incorrect Parameters, 3. Screen Lock, 4. Interruption of Bolus or Loading Dose Delivery, 5. Pump Displays Incorrect Bolus/Loading Dose, 6. Loading/Bolus Dose Below the Minimum Recommended Rate, 7. Motor Rate Error, 8. Incorrect Recall Last Settings, 9. Corrupt Configuration, 10. Auto Lock, 11. Toolbox Configuration Loading Dose Time Values. Smiths Medical corrected all issues included in this notification in previous software updates and the corrections were carried forward into all subsequent software releases. Please ensure you have the most recent Medfusion software (v6.0.0) installed on your pumps.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls 48IN (121.9CM) INJ LINE ROTATING ADAPT (900psi), List Number MX694R
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls 20IN (50.8CM) BRAIDED INJ LINE, ROTATING ADAPT, List Number MX682BR
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls HIGH PRESSURE ROTATOR WITH FEMALE LUER LOCK, List Number MX497HP
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls HIGH PRESSURE ROTATOR WITH MALE LUER LOCK, List Number MX496HP
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls Medex 700PSI STOPCOCK W/ROTATOR, List Number MX4331R
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls 6IN PRESSURE TUBING, List Number MX20617
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls 1050 STOPCOCK LEFT ROTATOR OFF HANDLE, List Number MX1431MRL
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
January 24, 2024Medical DevicesFDA
Smiths Medical Asd Inc recalls Medex LOGICAL CATH LAB KIT, List Number M20754
Hazard: There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.
November 8, 2023Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
Hazard: The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026).
November 1, 2023Medical DevicesFDA
Smiths Medical ASD Inc. recalls Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1
Hazard: The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under the FGSKU# 4599P-1 and Lot# 4331283. Tubes without caps have the potential to generate delayed or erroneous test results and/or exposure to infectious blood.