RecallWatchUSA

Home / Brands / Smiths Medical ASD

Smiths Medical ASD Recalls

293 recalls on record · July 18, 2012 to May 14, 2025 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

March 28, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls Medfusion Syringe Pump 4000 Series The Medfusion¿ Syringe Infusion Pumps are indicated for the following uses: 1) Adm
Hazard: Certain Medfusion Syringe Pump Models, Series 3100, 3500, and 4000, may not recognize or may misidentify loaded medication syringes. The inability of a pump to recognize a syringe (i.e. the size of the syringe is unknown to the pump) results in an inability to complete pump programming. Misidentification of a syringe is where the pump misinterprets the syringe size.
March 21, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls Power PORT-A-CATH II, TRAY POWER PAC II. Labeled as: a. VEN, PU, 2.6MM, UL 8.5 FR/INTRO, VALVED (1/EA); b. VEN
Hazard: Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
March 21, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU,
Hazard: Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.
February 21, 2018Medical DevicesFDA
Smiths Medical ASD, Inc. recalls Custom Made Tracheostomy Device The CADD Administration Sets with Flow Stop are designed to be used with a variety of
Hazard: Smiths Medical determined that a single custom Tracheostomy device had been shipped to the customer without required sterilization.
February 7, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls Graseby 3300 Pump and Graseby 3000 Series Pump Accessories Patient controlled syringe pump. Use for Infusion
Hazard: In certain cases, the lock and key supplied with Graseby 3300 pumps and accessories does not match the standard Graseby 3300 key.
January 10, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eurek
Hazard: Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.
January 10, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls Medfusion, Syringe Infusion Pump, Software Version V1.1, Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka
Hazard: Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.
January 10, 2018Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD-Legacy 1, Model 6400 Ambulatory Infusion Pump, REF 21-6400-51, PC 10008552-001, Smiths Medical ASD, Inc. Rx ONLY
Hazard: Certain CADD Legacy 1, Ambulatory Infusion Pumps, may contain non-matching serial numbers. The label attached to the back of the pump may list a serial number that does not match the correct serial number programmed into the software of the pump.
November 8, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD(TM) Yellow Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included., REF 21-7100-24,
Hazard: Smiths Medical became aware that certain Non Flow-Stop CADD(R) Medication Cassette Reservoirs may have been manufactured with an incorrect pressure plate.
November 8, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD(TM) Medication Cassette Reservoir with clamp and female Luer., Nonvented stopper included., 100 mL, For use with C
Hazard: Smiths Medical became aware that certain Non Flow-Stop CADD(R) Medication Cassette Reservoirs may have been manufactured with an incorrect pressure plate.
November 8, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD(TM) Medication Cassette Reservoir with clamp and female Luer. Nonvented stopper included., 50 mL, For use with CADD
Hazard: Smiths Medical became aware that certain Non Flow-Stop CADD(R) Medication Cassette Reservoirs may have been manufactured with an incorrect pressure plate.
October 4, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls UniPerc(R) Adjustable Flange Tracheostomy Tube kit, Reorder Number 100/897/080 CZ The UniPerc(R) Tracheostomy Tubes
Hazard: Smiths Medical became aware that one lot of the 8.0mm UniPerc(R) Adjustable Flange Tracheostomy Tube kit, product reorder number 100/897/080 CZ, lot number 3308971, contains an incorrect 9.0mm sized obturator instead of the correct 8.0mm sized obturator.
September 6, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD(R) Medication Cassette Reservoir with clamp and female Luer. Non vented stopper included, REF 21-7002-24, Rx only,
Hazard: The medication cassette reservoir, part number 21-7002-24, with lot number 16X659, may have been manufactured with the incorrect pressure plate and the tubing used on the cassette may have been routed incorrectly.
August 2, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls Smiths Medical CADD Blue-Striped IV Administration Set, REF 21-7036-01, Sterilized using ethylene oxide.
Hazard: The label of the CADD Blue-Striped Administrative Set is missing the expiration date, which is considered a regulatory compliance issue for Canada.
July 12, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls Jelco, Hypodermic needle Pro¿ needles 19G X1 1/2" 800/CA Model No. EL 1915 & 19G X 2" 600/CA Model No. SB 5021, Rx Only
Hazard: Smiths Medical became aware on January 11, 2017, that syringe assemblies provided by Nipro Medical Corporation contained pin holes in the side wall of the hub. The defect was not detected during Smiths Medical incoming inspection. The syringe assemblies were used to manufacture seven (7) finished good lots of Hypodermic Needle Pro¿ needles, Smiths Medical part numbers EL 1915 and SB5021.
June 21, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD¿ High Volume Administration Set with FlowStop, 21-7381-24. Product Usage: The CADD¿ High-Volume Administration
Hazard: One lot of CADD¿ High Volume Administration sets, part number 21-7381-24, lot number 26X035, incorrectly contained the IFU insert for part number 21-7357-24.
April 19, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls CADD¿ Administration Set, 21-7339-24, Rx Only, Sterile EO.
Hazard: Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD¿ Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on the IFU correctly identifies the product to include a 0.2 micron filter, the illustration on the IFU identifies the filter size as 2.0 microns.
March 22, 2017Medical DevicesFDA
Smiths Medical ASD Inc. recalls Bivona¿ TTS (Tight to Shaft) Pediatric Tracheostomy Tube, Model No. 67SP035, Reorder No. 67SP035.
Hazard: Smiths Medical became aware that one lot of the Bivona¿ 3.5mm TTS Cuffed Pediatric tracheostomy tubes was incorrectly labeled for neonatal use instead of correctly labeled for pediatric use.
March 22, 2017Medical DevicesFDA
Smiths Medical ASD, Inc. recalls CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,
Hazard: 20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.
February 15, 2017Medical DevicesFDA
Smiths Medical ASD, Inc. recalls HOTLINE Blood and Fluid Warmer, Mole l HL-90. Designed for use with the Hotline Fluid Warming Set (L-80, L-70 and L-70N
Hazard: HOTLINE Blood and Fluid Warmers included Greek and Danish language labeling instead of the required Finnish labeling.
February 1, 2017Medical DevicesFDA
Smiths Medical ASD, Inc. recalls NORMFlow H-110 Irrigation Fluid Warmer, Continental Europe, Model H-1100, Reorder H-1100-INT-230. Product Usage: NO
Hazard: Smiths Medical became aware that that one (1) NORMOFLO¿ H-1100 Irrigation Fluid Warmers was incorrectly labelled with Spanish language labels instead of English labels. Specifically five (5) labels were incorrect; the Operators Manual, Outer Base Box label, Pole Assembly Box label, and two (2) O-ring Lube labels.
December 28, 2016Medical DevicesFDA
Smiths Medical ASD Inc. recalls Acuvance¿2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance¿ 2 Safety Catheter is a Peripheral Intra
Hazard: Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves.
June 8, 2016Medical DevicesFDA
Smiths Medical ASD, Inc. recalls CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24
Hazard: Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.
October 28, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls PORT-A-CATH Plastic Hub Needles, 0.9mm (20G) x 19mm (3/4 in.) 90-degree. REF 21-2287-24 The Plastic hub PORT-A-CATH
Hazard: The label on the lid stock of the individual needle packages for a portion of Lot number 46262 contains the incorrect part number, part number description and product image. The part number should be 21-2287-24 Port-A-Cath Plastic Hub Bent Needle.
June 24, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls CADD¿- Solis Ambulatory Infusion Pump, Model 2111 & 2112, Version 1.0, Version 2.0 and Version 3.0 & Model 2120, Versio
Hazard: Smiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette Reservoir or Administration Set disposable to the pump. When binding occurs, it can prevent the key from fully rotating the Cassette/ Keypad Lock to the locked (e
June 17, 2015Medical DevicesFDA
Smiths Medical Asd Inc recalls Low Volume Extension Set, 72' w/Clamp FLL/MLL APV=1.3ml, STERILE, AND FLUID PATH NONPYROGENIC UNLESS PACKAGE IS DAMAGED
Hazard: Smiths Medical has become aware of an issue with the Low Volume Extension Set, (72'' Extension Set) where under-sized tubing was used in the production of 1 lot of product.
June 10, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls Smiths Medical, BCI Advisor Vital Signs Monitor, Product Reorder No 92M774325A.
Hazard: The Auxiliary Serial Input/ Output Modes on the Advisor do not function as expected, resulting in data formatting issues or incorrect data being transmitted in or out of the serial port into a data collection system.
March 25, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls CADD" Medication Cassette Reservoir, 50mL, REF 21-7001-24 & 21-7301-24, Sterile EO, Rx only.
Hazard: Smiths Medical has become aware of an issue with specific lots of 50mL CADD" Medication Cassette Reservoirs (Cassette). Some Cassettes may leak at the sealing area of the pump tube and medication bag. Smiths Medical has received no reports of serious injury or death related to this issue.
February 4, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls Portex Anaerobic Pulsator Arterial Blood Sampling Kit, with Liquid Sodium Heparin for Gases and Electrolytes. Reorder
Hazard: The outer box label is marked with the correct expiration date of 08/2016. The inner Tyvek Unit Label is marked with 08-2017 which exceeds the actual expiration date by one year.
January 7, 2015Medical DevicesFDA
Smiths Medical ASD, Inc. recalls Hypodermic Needle Pro¿ Edge" Safety Device with Syringe 20G x1 5mL Luer lock syringe; Reorder Number: 452010 Product
Hazard: Needle is not captured in the needle safety sheath in specific lots.