December 25, 2024Medical DevicesFDA
CooperSurgical, Inc. recalls Brand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Des
Hazard: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
December 25, 2024Medical DevicesFDA
CooperSurgical, Inc. recalls Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product De
Hazard: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
December 25, 2024Medical DevicesFDA
CooperSurgical, Inc. recalls Brand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An elect
Hazard: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
December 25, 2024Medical DevicesFDA
Philips North America Llc recalls Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Hazard: The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
December 25, 2024Medical DevicesFDA
Spacelabs Healthcare, Ltd. recalls Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch record
Hazard: Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
December 25, 2024Medical DevicesFDA
Spacelabs Healthcare, Ltd. recalls Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the pat
Hazard: Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
December 25, 2024Medical DevicesFDA
Spectranetics Corporation recalls Turbo Elite Laser Atherectomy Catheter Model Number / UDI-DI code / Product Description 414-151 / 989930000531 / 1.4 m
Hazard: Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.
December 25, 2024Medical DevicesFDA
Draeger Medical, Inc. recalls Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstatio
Hazard: The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
December 25, 2024Medical DevicesFDA
Draeger Medical, Inc. recalls Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation
Hazard: The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
December 25, 2024Medical DevicesFDA
Medtronic Neurosurgery recalls Exacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTA W/BRAID
Hazard: Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
December 25, 2024Medical DevicesFDA
Medtronic Neurosurgery recalls Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY V
Hazard: Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Optikon Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: OPT14-5011. Product Description: The Disposable Micro
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Nidek Co. Disposable Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: 18241-0979. Product Description: The Dis
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Walcott Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: RX14-5011. Product Description: The Disposable Micros
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Ambler Surgical Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: PD7226. The Disposable Microsurgical Bipolar
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Accutome/Microsurgical Disposable Bipolar Pencil 5 in. (12.7cm) 25 Gauge Straight. Model/Catalog Number: AB-7004 The
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Kirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight, Long. Model/Catalog Number: 14-5011L The D
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 25, 2024Medical DevicesFDA
Kirwan Surgical Products, LLC recalls Kirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight. Model/Catalog Number: 14-5011 The Disposab
Hazard: Potential for the outer stainless-steel tube to break or detach.
December 18, 2024Medical DevicesFDA
Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy S
Hazard: Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
December 18, 2024Medical DevicesFDA
Hermes Medical Solutions AB recalls Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x an
Hazard: Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
December 18, 2024Medical DevicesFDA
Diasorin Inc. recalls DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
Hazard: Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
December 18, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2.
Hazard: Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
December 18, 2024Medical DevicesFDA
Cardinal Health 200, LLC recalls Cardinal Health SMS Elastic Cuff Thumbhook Level 3 Isolation Gown, Full-Back, Product Codes AT6588-2XL, AT6588-BD, AT658
Hazard: Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
December 18, 2024Medical DevicesFDA
Cardinal Health 200, LLC recalls Convertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL
Hazard: Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
December 18, 2024Medical DevicesFDA
Beckman Coulter Inc. recalls Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determi
Hazard: The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
December 18, 2024Medical DevicesFDA
CareFusion 303, Inc. recalls BD Pyxis CII Safe ES, REF: 1116-00
Hazard: Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
December 18, 2024Medical DevicesFDA
CareFusion 303, Inc. recalls BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324
Hazard: Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
December 18, 2024Medical DevicesFDA
CareFusion 303, Inc. recalls BD Pyxis MedStation ES, REF: 323
Hazard: Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
December 18, 2024Medical DevicesFDA
Hologic, Inc recalls Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Nu
Hazard: IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
December 18, 2024Medical DevicesFDA
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV19
Hazard: During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.