January 21, 2026Medical DevicesFDA
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Sof
Hazard: increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
January 21, 2026Medical DevicesFDA
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 5081289 Softw
Hazard: increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
January 21, 2026Medical DevicesFDA
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software
Hazard: increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. M
Hazard: Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
January 21, 2026Medical DevicesFDA
Elekta, Inc. recalls MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)
Hazard: Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/20; Model Number: 722038;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/10; Model Number: 722029;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722028;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722027;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722026;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 OR Table; Model Number: 722023;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Model Number: 722013;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722012;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722011;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722010;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Model Number: 722008;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722006;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722005;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722003;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F; Model Number: 722002;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Product Name: Allura Xper FD10C; Model number: 722001;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
HANGZHOU BEVER MEDICAL DEVICES CO., LTD. recalls BD Ready-to-Use Hydrophilic Catheter 14 Fr (4.7 mm) Straight Tip, Coated, Female REF: RTU14F
Hazard: Labeling error, Incorrect expiration date
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-PLASTIC CRANIO
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: NEURO ANGIO PACK, M
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) ADMIN KIT, Medl
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medl
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) IV START KIT NS,
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) DRAWER 2 IV SUPP
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL P
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
January 21, 2026Medical DevicesFDA
Medline Industries, LP recalls Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: RR-LAP DONOR NEP AC
Hazard: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.