March 19, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
Hazard: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
March 19, 2025Medical DevicesFDA
GE Medical Systems, LLC recalls GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
Hazard: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
March 19, 2025Medical DevicesFDA
Smiths Medical ASD, Inc. recalls Portex Endotracheal Tube, Item Codes 100/105/025 100/105/030 100/105/035 100/111/020 100/111/025 100/111/030 100/111/035
Hazard: Affected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
March 12, 2025Medical DevicesFDA
Becton Dickinson & Co. recalls BD BBL TB Stain Kit K, Catalog (REF): 212522
Hazard: One component of the kit was shipped with an expiration date of 30Sep2024 while the Certificate of Analysis (COA) indicated an expiration date of 30Apr2025.
March 12, 2025Medical DevicesFDA
Integra LifeSciences Corp. recalls AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's Produ
Hazard: Possibility for the obturator to break (separate).
March 12, 2025Medical DevicesFDA
TAS Medical Inc recalls TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Hazard: Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
March 12, 2025Medical DevicesFDA
Alcon Research LLC recalls Alcon Custom Pak with Non-Latex Symbol, Pak: 12771-18, 18050-10, 18252-17, 12698-09, 15632-10, 3141-45, 12630-07, 18925-
Hazard: Customer-designed sterile surgical procedure packs contain latex components as specified; but, the content label sealed within the transparent sterile pouch contains the symbol for a latex-free product. If users or patients with latex sensitivity come into contact with latex, an allergic reaction may occur.
March 12, 2025Medical DevicesFDA
Murata Vios, Inc. recalls Vios Monitoring System Bedside Monitor Model BSM2050
Hazard: Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.
March 12, 2025Medical DevicesFDA
DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C
Hazard: Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
March 12, 2025Medical DevicesFDA
LUMENIS, LTD. recalls (1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, La
Hazard: The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls Osstem Implant System - Abutment: ET Multi AbutmentSize: 4.8D 5.0GH - Abutment is intended for use with a dental implan
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls ET Rigid Abutment (Mini) Size: 4.5D 2.0G/H 4.0H- Intended for use with a dental implant to provide support for prostheti
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls Osstem Implant System - Abutment ET Multi Abutment Size: 4.8D 3.0G/H- Abutment is intended for use with a dental implan
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls ET Rigid Abutment (Mini) SIze: 4.5D 4.0G/H 4.0H - Intended for use with a dental implant to provide support for prosthet
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls ET Rigid Abutment (Mini) SIze: 4.5D 3.0G/H 7.0H - Intended for use with a dental implant to provide support for prosthet
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Hiossen Inc. recalls ET Rigid Abutment (Mini) Size 4.5D 1.0G/H 5.0H - Intended for use with a dental implant to provide support for prosthet
Hazard: Patient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
March 12, 2025Medical DevicesFDA
Telcare, LLC recalls Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;
Hazard: Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.
March 12, 2025Medical DevicesFDA
Kung Shin Plastics Co. Ltd. recalls CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
Hazard: The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
March 12, 2025Medical DevicesFDA
Kung Shin Plastics Co. Ltd. recalls CircuitGuard ThermoFlo Filter -Combined Heat and Moisture Exchanger, Bacterial Viral Filter w O2 Port and Elbow (Model/C
Hazard: The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
March 12, 2025Medical DevicesFDA
MIM Software Inc recalls MIM software; System, Image Processing, Radiological
Hazard: In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 740G (MMT-1811, MMT-1812, MMT-1861, MMT-1862)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 720G (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 700G (MMT-1801, MMT-1805, MMT-1850, MMT-1851)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 640G (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls MiniMed insulin pump, REF: 620G (MMT-1750)
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
March 12, 2025Medical DevicesFDA
Medtronic MiniMed, Inc. recalls Paradigm insulin pump, REF: MMT-754
Hazard: All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.